Caractéristiques générales :
- Décharge automatique de la tension
- Fonction Max/Min
- Thermocouple Type K
- Arrêt automatique
- Grand écran LCD rétroéclairé
- CAT III 1000V, CAT IV 600V

- IP67
- Dimensions : 220 x 96.5 x 60.5mm
- Poids : 631g
- Complet avec : Mallette rigide, sonde de température, câble USB, logiciel, manuel

| TENSION CC | |||||
| GAMME | RÉSOLUTION | PRÉCISION | |||
| 400mV | 0.01mV | ±(0.06% rdg +4 dgt) | |||
| 4V | 0.0001V | ||||
| 40V | 0.001V | ||||
| 400V | 0.01V | ||||
| 1000V | 0.1V | ±(0.1% rdg +5 dgt) | |||
| TENSION CA | |||||
| GAMME | RÉSOLUTION | PRÉCISION | |||
| de 50 à 1000Hz | |||||
| 400mV | 0.1mV | ±(1.0% rdg +7 dgt) | |||
| 4V | 0.001V | ||||
| 40V | 0.01V | ||||
| 400V | 0.1V | ||||
| 1000V | 1V | ±(1.0% rdg +5 dgt) | |||
| TENSION CA+CC | |||||
| GAMME | RÉSOLUTION | PRÉCISION | |||
| 400mV | 0.1mV | ±(1.0% rdg +7 dgt) (50/60HZ) | |||
| 4V | 0.001V | ||||
| 40V | 0.01V | ||||
| 400V | 0.1V | ||||
| 1000V | 1V | ||||
| Note : toutes les plages de tension CA se réfèrent à l'intervalle de la plage comprise entre 5% et 100% | |||||
| COURANT CC | |||||
| GAMME | RÉSOLUTION | PRÉCISION | |||
| 400µA | 0.01µA | ±(1.0% rdg +3 dgt) | |||
| 4000µA | 0.1µA | ||||
| 40mA | 0.001mA | ||||
| 400mA | 0.01mA | ||||
| 10A | 0.001A | ||||
| (20A : maximum 30 sec. avec précision réduite) | |||||
| COURANT CA (CA+CC) | |||||
| GAMME | RÉSOLUTION | PRÉCISION | |||
| de 50 à 1000Hz | |||||
| 400µA | 0.1µA | ±(1.5% rdg +7 dgt) | |||
| 4000µA | 1µA | ||||
| 40mA | 0.01mA | ||||
| 400mA | 0.1mA | ||||
| 10A | 0.01A | ||||
| COURANT CA+CC | |||||
| GAMME | RÉSOLUTION | PRÉCISION | |||
| 400µA | 0.1µA | ±(1.5% rdg +7 dgt) | |||
| 4000µA | 1µA | ||||
| 40mA | 0.01mA | ||||
| 400mA | 0.1mA | ||||
| 10A | 0.01A | ||||
| (20A : maximum 30 sec. avec précision réduite) Note : toutes les plages de tension CA se réfèrent à l'intervalle de la plage comprise entre 5% et 100% | |||||
| RÉSISTANCE | |||||
| GAMME | RÉSOLUTION | PRÉCISION | |||
| 400 Ohm | 0.1 Ohm | ±(0.3% rdg +9 dgt) | |||
| 4 kOhm | 0.0001 kOhm | ±(0.3% rdg +4 dgt) | |||
| 40 kOhm | 0.001 kOhm | ||||
| 400 kOhm | 0.01 kOhm | ||||
| 4 MOhm | 0.001 MOhm | ||||
| 40 MOhm | 0.001 MOhm | ±(2.0% rdg +10 dgt) | |||
| CAPACITÉ | |||||
| GAMME | RÉSOLUTION | PRÉCISION | |||
| 40 nF | 0.001 nF | ±(3.5% rdg +40 dgt) | |||
| 400 nF | 0.01 nF | ±(3.5% lecture +10 digits) | |||
| 4 µF | 0.0001 µF | ||||
| 40 µF | 0.001 µF | ||||
| 400 µF | 0.01 µF | ||||
| 4000 µF | 0.1 µF | ±(5% lecture +10 digits) | |||
| 40 mF | 0.001 mF | ||||
| FRÉQUENCE (électronique) | |||||
| GAMME | RÉSOLUTION | PRÉCISION | |||
| 40 Hz | 0.001 Hz | ±(0.1% lecture +1 digit) | |||
| 400 Hz | 0.01 Hz | ||||
| 4 kHz | 0.0001 kHz | ||||
| 40 kHz | 0.001 kHz | ||||
| 400 kHz | 0.01 kHz | ||||
| 4 MHz | 0.0001 MHz | ||||
| 40 MHz | 0.001 MHz | ||||
| 100 MHz | 0.01 MHz | Non spécifié | |||
|
Sensibilité : 0.8V rms min. de 20% à 80% rapport cyclique et <100kHz ; 5Vrms min de 20% à 80% rapport cyclique et > 100kHz. | |||||
| FRÉQUENCE (électrique) | |||||
| GAMME | RÉSOLUTION | PRÉCISION | |||
| 40.00 Hz - 10 kHz | 0.01 Hz - 0.001 kHz | ±(0.5% lecture) | |||
| Sensibilité : 1V rms | |||||
| Rapport cyclique | |||||
| GAMME | RÉSOLUTION | PRÉCISION | |||
| 0.1 - 99.90% | 0.01% | ±(1.2% lecture +2 digits) | |||
|
Largeur d'impulsion : 100 µs - 100ms Fréquence : de 5 Hz à 150 kHz | |||||
| TEMPÉRATURE (type K) | |||||
| GAMME | RÉSOLUTION | PRÉCISION | |||
| de -50 à 1000°C | 0.1°C | ±(1.0% lecture +2.5°C) | |||
| de -58 à 1832°F | 0.1°F | ±(1.0% lecture +4.5°F) (précision de la sonde non incluse) | |||
| 4-20mA% | |||||
| GAMME | RÉSOLUTION | PRÉCISION | |||
| de -25 à 125% | 0.01% | ±50 digits | |||
| 0mA=-25%, 4mA=0%, 20mA=100%, 24mA=125% | |||||
| MÉGOHMS | |||||
| Tension de la borne | GAMME | RÉSOLUTION | PRÉCISION | Courant d'essai | Courant de court-circuit |
| 125V (0% - +10%) | 0.125 - 4.000 MOhm | 0.001 MOhm | +(2%+10) | 1mA avec charge 125 kOhm | < 1mA |
| 4.001 - 40.00 MOhm | 0.01 MOhm | +(2%+10) | |||
| 40.01 - 400.0 MOhm | 0.1 MOhm | ||||